RESEARCH ON THE PARAMETERS OF A PASSIVE (INTEGRAL) DOSIMETRIC SYSTEM FOR NATURAL RADIATION BACKGROUND MEASUREMENT
Tsvetelin P. Tsrunchev, Daniela P. Draganova, Georgi E. Apostolov
Pages: 40-47
Received: 30 Apr 2025
Published: 3 Dec 2025
DOI: 10.62991/ES1996606656
Views: 427
Downloads: 35
Abstract: The TLD RADOS-2000-A dosimetric system is designed for measuring the ambient dose H*(10) related to environmental and workplace radiation. The system includes environmental thermoluminescent dosimeters and a thermoluminescent reader. Each dosimetric system must be calibrated before commissioning to establish the relationship between irradiation parameters and the measured controlled parameter. The relationship between the ambient dose H*(10) and the luminescence read by the thermoluminescent reader is applied for measurement. This dependency is influenced by the composition of the dosimeter tablets and the production conditions. Thus, batch homogeneity and dosimeter separation must be investigated. This article describes and documents the initial procedures in commissioning a dosimetric system—separation and calibration of the dosimeters. The methods applied are based on international standards and guidelines, as well as on the long-standing experience of NCRRP in using thermoluminescent dosimeters for measuring individual equivalent doses in personal dosimetric control. Various factors influence the readings of dosimetric calibration measurements. One significant factor is the angular dependence of the radiation incident on the dosimeters. At this stage, the effect was studied at incident angles of α = 0°, ±60°, and ±90°. The studies and calibrations were carried out using the G10 Gamma Beam Irradiator System, Hopewell Designs Inc., loaded with a 137Cs source. The system includes a positioning system, a reference measurement system (electrometer and ionization chamber). Using this system ensures high calibration accuracy and traceability of dosimetric measurements.
Keywords: radiation, dosimetry, tld, background, calibration, measurements
Cite this article: Tsvetelin P. Tsrunchev, Daniela P. Draganova, Georgi E. Apostolov. RESEARCH ON THE PARAMETERS OF A PASSIVE (INTEGRAL) DOSIMETRIC SYSTEM FOR NATURAL RADIATION BACKGROUND MEASUREMENT. Journal of International Scientific Publications: Ecology & Safety 19, 40-47 (2025). https://doi.org/10.62991/ES1996606656
Back to the contents of the volume

Submit Feedback

We value your input! Use this form to report any concerns or provide feedback on our published articles. All submissions will be kept confidential.

By using this site you agree to our Privacy Policy and Terms of Use. We use cookies, including for analytics, personalisation, and ads.